Autotest sequences
41
Insulation resistance measurement
This test is offered if it is applicable according to the VDE Organizer setting.
The measurement is described in chapter
5.2.3 Insulation resistance
.
Figure 6.5: VDE Organizer – Insulation resistance starting screen
Options in insulation resistance starting screen:
500 V
Standard test voltage
250 V
To be set if overvoltage protection devices are installed or SELV/ PELV
protection measures.
Figure 6.6: VDE Organizer – Insulation resistance result screen
There are no special options to be set in the insulation resistance result screen.
Substitute leakage measurement
This test is offered if it is applicable according to the VDE Organizer setting.
The measurement is described in chapter
5.2.5 Substitute leakage
.
Figure 6.7: VDE Organizer – Substitute leakage starting screens
Options if
Appliances with heating elements
is set:
/
Sets power of heating elements.
Note:
The leakage current limit is automatically set on base of set appliance power.
There are no special options to be set if other Appliance type is set.
Figure 6.8: VDE Organizer – Substitute leakage result screen
There are no special options to be set in the Substitute leakage result screen.