MI 3311 GammaGT
Autotest sequences
42
Insulation resistance - P measurement
This test is offered if it is applicable according to the VDE Organizer setting.
The measurement is described in chapter
5.2.4 Insulation resistance - P
.
Figure 6.9: VDE Organizer autotest – Insulation resistance - P starting screen
Options in Insulation resistance - P starting screen:
500 V
Standard test voltage
250 V
To be set if overvoltage protection devices are installed or SELV/ PELV
protection measures.
Figure 6.10: VDE Organizer – Insulation resistance - P result screen
Options in Insulation resistance - P result screen:
NEXT
Proceeds with the next measurement.
REPEAT
Repeats the test (use in case of multiple isolated/SELV/PELV
accessible points). Lowest result will be stored.
Substitute Leakage - P current measurement
This test is offered out if it is applicable according to the VDE Organizer setting.
The measurement is described in chapter
5.2.6 Substitute leakage - P.
Figure 6.11: VDE Organizer – Substitute leakage - P starting screen
There are no special options in the Substitute leakage - P starting screen.
Figure 6.12: VDE Organizer – Substitute leakage - P result screen
Options in Substitute leakage - P result screen:
NEXT
Proceeds with the next measurement.
REPEAT
Repeats the test (use in case of multiple isolated/SELV/PELV
accessible points). Highest result will be stored.