MI 3311 GammaGT
Autotest sequences
44
6.2 Custom autotest
In the autotest custom menu user-defined autotest procedures can be performed via the
PC SW PAT Link PRO. Up to 50 custom autotest sequences can be pre-programmed in
this autotest mode.
Commonly used pre-programmed autotest sequences are added to the list by default.
The custom sequences can be also uploaded from the PC software PAT Link PRO.
Refer to chapter
8 Communication
for more information.
New custom sequences can be also uploaded from the VDE Organizer.
Refer to chapter
7 Working with autotest results f
or more information.
The pre-programmed sequences can be restored to default settings by selecting
Initial
settings
in
Setup menu
.
Select Custom Autotest function in Main menu.
Figure 6.14: Custom autotest menu
Keys:
/
Selects the custom autotest.
START
Starts the selected custom autotest. See chapter
6.5 Carrying out (Code,
Simple and Custom) autotest sequences
.
ESC
Returns to Main menu.
Note:
If more than 50 autotests are saved, »Out of custom autotest memory« message is
displayed.
6.3 Simple test
Simple test sequences are commonly used pre-programed autotest sequences with
possibility of fast testing. Fast testing mode can be enabled in Test speed setup function in
Setup menu. Refer to chapter
4.8.5 Test speed setup
for more information.
Select Simple test function in Main menu.
Figure 6.15: Simple test menu
Keys:
/
Selects the test sequence from the list
START
Starts the selected test. See chapter
6.5 Carrying out (Code, Simple and
Custom) autotest sequences
.
ESC
Returns to Main menu.
Note:
Limits of the CLASS I, II and III tests are shown in
Appendix D