MI 3394 CE MultiTesterXA
Single
tests
20
Test parameters
Starting DC test voltage
U start [500 V ... 6000 V in steps of 50 V]
DC test voltage
U test [500 V ... 6000 V in steps of 50 V]
Duration of starting voltage
t start [1 s ... 120 s ]
Duration of ramp
t ramp [2 s ... 60 s ]
Duration of test voltage
t end [Off, 1 s ... 120 s ]
Test limits
High limit (I)
H limit [0.05 mA ... 10.0 mA ]
Low limit (I) L limit [Off, 0.05 mA ... 10.0 mA]
Test circuit
Figure 4.16: HV DC programmable test
HV DC programmable test procedure
Select the
HV DC
programmable
function.
Set test parameters / limits.
Connect HV test leads to HV(~,+) and HV(~,-) terminals on the instrument.
Connect HV test leads to device under test.
Start measurement.
Measurement can be stopped manually or by timer.
Save results (optional).
Figure 4.17: Examples of HV DC programmable test results