MI 3394 CE MultiTesterXA
Single
tests
18
Test results / sub-results
I ................. test current
U ............... measured test voltage
Ir ............... resistive portion of test current
Ic ............... capacitive portion of test current
Test parameters
Starting AC test voltage
U start [100 V ... 5000 V in steps of 10 V]
AC test voltage
U test [100 V ... 5000 V in steps of 10 V]
Duration of starting voltage
t start [1 s ... 120 s ]
Duration of ramp
t ramp [2 s ... 60 s ]
Duration of test voltage
t end [Off, 1 s ... 120 s ]
Test limits
High limit (I)
H limit [0.5 mA ... 100 mA ]
Low limit (I) L limit [Off, 0.5 mA ... 100 mA]
Test circuit
Figure 4.13: HV AC programmable test