MI 3394 CE MultiTesterXA
Single
tests
13
Figure 4.3: Measurement of Continuity P/S - PE
Continuity measurement procedure
Select the
Continuity
function.
Set test parameters / limits.
Connect test leads to C1, P1, P2 and C2 terminals on the instrument (4 wire), or connect
test lead to P/S terminal (2 wire measurement P/S – PE).
Connect test leads to device under test.
Start measurement.
Measurement can be stopped manually or by timer.
Save results (optional).
Figure 4.4: Examples of Continuity measurement results
4.1.2 HV
AC
IMPORTANT SAFETY NOTE
Refer to chapter
1.1 Warnings and notes
for more information regarding safe use of the
instrument.