MI 3394 CE MultiTesterXA
Single
tests
30
Test limits
H Limit (Itou)
H limit [Off, 0.25 mA ... 15.0 mA ]
L Limit (Itou)
L limit [Off, 0.25 mA ... 15.0 mA ]
Test circuit
Figure 4.35: Measurement of Touch Leakage current
Touch Leakage measurement procedure
Select the
Touch Leakage
function.
Set test parameters / limits.
Connect device under test to mains test socket. Connect test lead to P/S terminal on the
instrument and on device under test.
Start measurement.
Measurement can be stopped manually or by timer.
Save results (optional).
Figure 4.36: Examples of Touch Leakage measurement results