MI 3394 CE MultiTesterXA
Single
tests
14
Figure 4.5: HV AC test menu
Test results / sub-results
I ................. test current
U ............... measured a.c. test voltage
Ir ............... resistive portion of test current
Ic ............... capacitive portion of test current
Test parameters
AC test voltage
U test [100 V ... 5000 V in steps of 10 V]
Duration
t end [Off, 1 s ... 120 s]
Test limits
High limit (I)
H limit [0.5 mA ... 100 mA ]
Low limit (I) L limit [Off, 0.5 mA ... 100 mA]
Test circuit
Figure 4.6: HV AC measurement
HV AC measurement procedure
Select the
HV AC
function.
Set test parameters / limits.
Connect HV test leads to HV(~,+) and HV(~,-) terminals on the instrument.
Connect HV test leads to device under test.