MI 3394 CE MultiTesterXA
Single
tests
23
Figure 4.21: Measurement of Riso, Riso-S (socket)
RISO measurement procedure
Select the
Riso
function.
Set test parameters / limits.
Connect test leads to ISO(+), ISO(-) terminals on the instrument, then connect test leads to
device under test, or
Connect device to mains test socket. For Riso-S test, additionally connect test lead to P/S
terminal on instrument, and then connect test lead to device.
Start measurement.
Measurement can be stopped manually or by timer.
Save results (optional).
Figure 4.22: Examples of Insulation resistance measurement results
Note:
When P/S probe is connected during the Riso measurement, then the current through it
is also considered.