![Meggitt vibro-meter SpeedSys300 ODS301 User Manual Download Page 34](http://html1.mh-extra.com/html/meggitt/vibro-meter-speedsys300-ods301/vibro-meter-speedsys300-ods301_user-manual_1765658034.webp)
USER MANUAL
Document reference MAODS301/E
SpeedSys300 ODS301 overspeed detection system
Edition 4 - May 2022
2 - 10
SpeedSys300 ODS301 module connectors
OVERVIEW OF THE SPEEDSYS300 ODS301 MODULE HARDWARE
Connector pinouts (definition of terminals) for the 8 connectors on the bottom of the module
are given in
to
Table 2-6:
SpeedSys300 ODS301 module SAFETY ANALOG / FREQ OUT connector pinouts
SpeedSys300 ODS301 SAFETY ANALOG / FREQ OUT connector
Label
Pin
Name /
function
Direction
Definition
S
A
FE
T
Y
ANA
L
OG /
FRE
Q
OUT
A13
MA+
B
Safety analog (4 to 20 mA current) output: +
A14
MA−
B
Safety analog (4 to 20 mA current) output: −
A15
FREQ+
O
Frequency output: Signal
A16
FREQ−
B
Frequency output: GND
Notes
B = bidirectional, GND = ground, O = output.
Table 2-7:
SpeedSys300 ODS301 module POWER PSU 1 connector pinouts
SpeedSys300 ODS301 POWER PSU 1 connector
Label
Pin
Name /
function
Direction
Definition
POW
E
R
PS
U
1
A17
V+
I
Power supply 1: +
A18
V0
B
Power supply 1: − / GND
A19
IE
B
Instrument earth / GND
A20
IE
B
Instrument earth / GND
Notes
B = bidirectional, GND = ground, I = input, IE = Instrument earth.
Table 2-8:
SpeedSys300 ODS301 module POWER PSU 2 connector pinouts
SpeedSys300 ODS301 POWER PSU 2 connector
Label
Pin
Name /
function
Direction
Definition
PO
W
E
R
PS
U
2
A21
V+
I
Power supply 2: +
A22
V0
B
Power supply 2: − / GND
A23
IE
B
Instrument earth / GND
A24
IE
B
Instrument earth / GND
Notes
B = bidirectional, GND = ground, I = input, IE = Instrument earth.
Summary of Contents for vibro-meter SpeedSys300 ODS301