Troubleshooting Guide
15 - 9
Suggested Troubleshooting Sequence
When a condition occurs, observe the Sentinel message code or
symptom, and determine if the condition is being caused by the Con-
trol or Injection group of circuits. Isolate the condition to a board or
boards, replace the board(s) or troubleshoot to component level. See
Figure 15.2 below.
Figure 15.2: Suggested Troubleshooting Sequence
1. Condition
2. Message or Symptom
3. Circuit Groups
Control Group
CPU Card
I/O Card
Servo Control Card
(Digital Portion)
PPI Card
Display Card
Injection Group
Servo Control Card
(Analog Portion)
Power Drive Card
Mechanical Stop
4. Circuit Board
5. Circuit
Summary of Contents for ProVis Mark V
Page 1: ...Service Manual KMP 870 98411 T 205 Rev D...
Page 2: ......
Page 3: ......
Page 4: ......
Page 15: ...Operating Guide 3 1 3 Operating Guide...
Page 16: ...Mark V ProVis Injection System 3 2...
Page 30: ...Mark V ProVis Injection System 7 4 Figure 7 1 Block Diagram Servo Control Card...
Page 31: ...Servo Control Card 7 5...
Page 37: ...Mark V ProVis Injection System 9 4 Figure 9 1 Block Diagram Control Panel Digital Logic...
Page 39: ...Mark V ProVis Injection System 9 6 Figure 9 2 Block Diagram Control Panel DC to DC Converter...
Page 49: ...Mark V ProVis Injection System 10 10 NOTES...
Page 60: ...Injector Head 13 7 Figure 13 1 Block Diagram Head Card...
Page 61: ...Mark V ProVis Injection System 13 8 NOTES...
Page 64: ...Imaging System Interface 14 3 Figure 14 1 Connector Locations...
Page 67: ...Mark V ProVis Injection System 14 6 Figure 14 2 Block Diagram Imaging System Interface...
Page 71: ...Mark V ProVis Injection System 14 10 NOTES...