Mark V ProVis Injection System
15 - 30
VERIFY PRESSURE
An attempt was made to arm the injector without a Pressure value
entered. Press Delta 6, then enter a Pressure limit value.
VERIFY VOLUME
An attempt was made to arm the injector without a Volume value
entered or calculated. Enter a Volume value, or reenter the Flow Rate
and Injection Duration.
VOLUME GREATER THAN SYRINGE SIZE
A Volume was entered that exceeds the size of the installed syringe.
Lower the Volume amount, or change to a larger syringe size.
XX IS USED, CLEAR IT_YES/NO?
A program is already stored in the PPI storage location selected.
Press YES to erase the stored program and replace with a new pro-
gram in the same location. Press NO to save the existing program.
Re-enter a different storage location number for the new program.
XXX.X VOLUME REMAINS. ARM? YES/NO
The SINGLE arm key has been pressed in an attempt to arm the
injector when insufficient volume remains in the syringe to deliver the
program volume. The option is given to the user to either inject the
remaining insufficient volume, or to temporarily abandon arming until
program volume can be reprogrammed or syringe size can be
changed.
Press the YES key to complete the arming sequence
Press the NO key to abandon the arming sequence.
XXX.X VOLUME IS INSUFFICIENT TO ARM
The MULTIple arm key has been pressed in an attempt to arm the
injector when insufficient volume remains in the syringe to deliver one
bolus of the program volume. This message will be displayed for
approximately 3 seconds and indicates to the user that the current
remaining volume is insufficient to perform multiple injections accord-
ing to the program volume selected.
60/150 SYRINGE SYSTEM INSTALLED
The user attempted a size change to a 200 ml syringe from the Recall
50 utility program when the system software/hardware is configured
for the 60/150 syringe system. Use Recall 65 to access the 150/200
ml syringe system and select the 200 ml syringe size.
NOTE
: The Mark V
ProVis
Injector should not be configured
for 200 ml syringe operation.
Summary of Contents for ProVis Mark V
Page 1: ...Service Manual KMP 870 98411 T 205 Rev D...
Page 2: ......
Page 3: ......
Page 4: ......
Page 15: ...Operating Guide 3 1 3 Operating Guide...
Page 16: ...Mark V ProVis Injection System 3 2...
Page 30: ...Mark V ProVis Injection System 7 4 Figure 7 1 Block Diagram Servo Control Card...
Page 31: ...Servo Control Card 7 5...
Page 37: ...Mark V ProVis Injection System 9 4 Figure 9 1 Block Diagram Control Panel Digital Logic...
Page 39: ...Mark V ProVis Injection System 9 6 Figure 9 2 Block Diagram Control Panel DC to DC Converter...
Page 49: ...Mark V ProVis Injection System 10 10 NOTES...
Page 60: ...Injector Head 13 7 Figure 13 1 Block Diagram Head Card...
Page 61: ...Mark V ProVis Injection System 13 8 NOTES...
Page 64: ...Imaging System Interface 14 3 Figure 14 1 Connector Locations...
Page 67: ...Mark V ProVis Injection System 14 6 Figure 14 2 Block Diagram Imaging System Interface...
Page 71: ...Mark V ProVis Injection System 14 10 NOTES...