Mark V ProVis Injection System
15 - 22
MINIMUM FLOW RATE IS .3
A Flow Rate value less than .3 was selected. Enter a Flow Rate value
of at least .3 ml/sec/min/hr.
MISSING DATA CODE ON RCV
The RS-232C interface (from the Control Panel) is sending an error
message.
Procedure:
1. Remove power from the injector for ten seconds, then
reapply power and retry. There may have been transient
interference on the lines.
2. If the condition persists, replace the PPI card.
MOVE PLUNGER FWD 2 ML, THEN REARM
When power is first applied, this message is displayed if the operator
fails to advance the injector head plunger at least 2 ml prior to arming.
To correct:
1. Advance the plunger forward (or reverse) at least 2 ml.
Verification is indicated by 5 short beeps.
2. Rearm the injector.
NOTE
: If the plunger is in full reverse, pressing reverse load
will prompt the Sentinel message PLUNGER POT
MECHANICAL FAILURE. To correct, remove injector
power, wait 10 seconds, reapply power, then advance
the plunger forward.
NOTE
: If power is removed and reapplied at any time, the
plunger must be advanced forward 2 ml before arming.
**NMI - WATCHDOG 1 FAILED
Watchdog 1 on the CPU Card is not being triggered, sending an inter-
rupt signal to the CPU.
Procedure:
1. Remove power from the injector for 10 seconds, reapply
power and retry.
2. If the condition persists, replace the CPU Card.
NOT ENOUGH VOLUME FOR INJECTION
This message appears during MULTIple injections. Insufficient Vol-
ume remains in the syringe for additional injections. Select a lower
Volume value, or load additional media into the syringe.
Summary of Contents for ProVis Mark V
Page 1: ...Service Manual KMP 870 98411 T 205 Rev D...
Page 2: ......
Page 3: ......
Page 4: ......
Page 15: ...Operating Guide 3 1 3 Operating Guide...
Page 16: ...Mark V ProVis Injection System 3 2...
Page 30: ...Mark V ProVis Injection System 7 4 Figure 7 1 Block Diagram Servo Control Card...
Page 31: ...Servo Control Card 7 5...
Page 37: ...Mark V ProVis Injection System 9 4 Figure 9 1 Block Diagram Control Panel Digital Logic...
Page 39: ...Mark V ProVis Injection System 9 6 Figure 9 2 Block Diagram Control Panel DC to DC Converter...
Page 49: ...Mark V ProVis Injection System 10 10 NOTES...
Page 60: ...Injector Head 13 7 Figure 13 1 Block Diagram Head Card...
Page 61: ...Mark V ProVis Injection System 13 8 NOTES...
Page 64: ...Imaging System Interface 14 3 Figure 14 1 Connector Locations...
Page 67: ...Mark V ProVis Injection System 14 6 Figure 14 2 Block Diagram Imaging System Interface...
Page 71: ...Mark V ProVis Injection System 14 10 NOTES...