Mark V ProVis Injection System
15 - 20
MAXIMUM FLOW RATE IS 20
A Flow Rate above 20 was entered while in the 60 ml syringe system.
Press RESET, then enter a Flow Rate less than 20.
MAXIMUM FLOW RATE IS 50 ML/SEC
Injector is in the ML/SEC Flow scale, and a Flow Rate above 50 ml/
sec was entered. Press RESET, then enter a value no greater than 50
ml/sec.
MAXIMUM RISE/FALL IS 9.9 SECONDS
A Rise/Fall time greater than 9.9 seconds was entered. Press RESET,
then enter a value no greater than 9.9 seconds.
MAX PRESSURE IN ML/MIN/HR IS XXXX
A pressure limit was entered that is greater than the maximum
allowed in any given flow scale or syringe size. Enter a pressure limit
value no greater than the respective maximum.
MAX PRESSURE IN ML/SEC IS XXXX
A pressure limit which exceeds the maximum allowed in the selected
syringe system and flow scale. Enter a pressure limit no greater than
the respective maximum. (See acceptable pressure value ranges).
MAX PRESSURE IN ML/SEC IS 82
A pressure limit above 82 atü (atmospheres) was entered while using
the ml/sec Flow scale. This message appears when using 150 ml
syringes and the atü Pressure scale. Press RESET, then enter a value
no greater than 82 atü.
MAX PRESSURE IN ML/SEC IS 84
A pressure limit above 84 kg/sq cm was entered while using the ml/
sec Flow scale. This message appears when using 150 ml syringes
and the kg (kg/sq cm) Pressure scale. Press RESET, then enter a
value no greater than 84 kg/sq cm.
Minimum and Maximum Pressure Values
60 ml Syringe
150 and 200 ml Syringe
psi
kg/cm
2
kPa
atü
psi
kg/cm
2
kPa
atü
100
7
689
6
100
7
689
6
ml / SEC
1100
77
7584
75
ml / SEC
1200*
84*
8274*
82*
100
7
689
6
100
7
689
6
ml / MIN / HR
700
49
4826
48
ml / MIN / HR
700
49
4826
48
* Values shown are for the 150 ml syringe
ml/SEC values for 200 ml syringe are: 1000 70 6895 68
Summary of Contents for ProVis Mark V
Page 1: ...Service Manual KMP 870 98411 T 205 Rev D...
Page 2: ......
Page 3: ......
Page 4: ......
Page 15: ...Operating Guide 3 1 3 Operating Guide...
Page 16: ...Mark V ProVis Injection System 3 2...
Page 30: ...Mark V ProVis Injection System 7 4 Figure 7 1 Block Diagram Servo Control Card...
Page 31: ...Servo Control Card 7 5...
Page 37: ...Mark V ProVis Injection System 9 4 Figure 9 1 Block Diagram Control Panel Digital Logic...
Page 39: ...Mark V ProVis Injection System 9 6 Figure 9 2 Block Diagram Control Panel DC to DC Converter...
Page 49: ...Mark V ProVis Injection System 10 10 NOTES...
Page 60: ...Injector Head 13 7 Figure 13 1 Block Diagram Head Card...
Page 61: ...Mark V ProVis Injection System 13 8 NOTES...
Page 64: ...Imaging System Interface 14 3 Figure 14 1 Connector Locations...
Page 67: ...Mark V ProVis Injection System 14 6 Figure 14 2 Block Diagram Imaging System Interface...
Page 71: ...Mark V ProVis Injection System 14 10 NOTES...