Disassembly and Reassembly
16 - 15
Injector Head
- Syringe Heat Maintainer
(continued)
3. Install the wires/pins into the J29 connector housing (white -Pin 1,
black -Pin 2). Ensure that the tangs lock into place when inserting
pins into the connector housing. After installing the wires/pins, gently
tug on each wire to verify that the pins are secured in the connector.
4. Reconnect the P29/J29 connectors and the injector head to the
injector unit, and apply unit power. The heater should feel warm
within a few minutes. If the over-temperature light illuminates, place
the syringe heat maintainer on a pressure jacket; the light should
extinguish within a few minutes. If the over temperature light remains
lit, the assembly is defective.
5. After verifying correct operation, remove unit power. Reinstall the
bottom cover and retainer screws.
Summary of Contents for ProVis Mark V
Page 1: ...Service Manual KMP 870 98411 T 205 Rev D...
Page 2: ......
Page 3: ......
Page 4: ......
Page 15: ...Operating Guide 3 1 3 Operating Guide...
Page 16: ...Mark V ProVis Injection System 3 2...
Page 30: ...Mark V ProVis Injection System 7 4 Figure 7 1 Block Diagram Servo Control Card...
Page 31: ...Servo Control Card 7 5...
Page 37: ...Mark V ProVis Injection System 9 4 Figure 9 1 Block Diagram Control Panel Digital Logic...
Page 39: ...Mark V ProVis Injection System 9 6 Figure 9 2 Block Diagram Control Panel DC to DC Converter...
Page 49: ...Mark V ProVis Injection System 10 10 NOTES...
Page 60: ...Injector Head 13 7 Figure 13 1 Block Diagram Head Card...
Page 61: ...Mark V ProVis Injection System 13 8 NOTES...
Page 64: ...Imaging System Interface 14 3 Figure 14 1 Connector Locations...
Page 67: ...Mark V ProVis Injection System 14 6 Figure 14 2 Block Diagram Imaging System Interface...
Page 71: ...Mark V ProVis Injection System 14 10 NOTES...