Mark V ProVis Injection System
7 - 2
PIO3 include: Flow Control signals, the Safe Relay Enable, and the
System Override signal.
System Monitor
Flip-Flop
The flip-flop (U17) sends alarm signals to disarm the injector, if any of
the following input conditions occur: WATCHDOG 2, OPLIM (over-
pressure signal), or AUX MONITOR (power drive failure).
Command, Position,
and Signal DACs
When enabled and timed by the DAC decoder U14, one of the DACs
(U21-U24) will read a byte of data bus information. The DAC then
converts the byte to an analog voltage between zero and 10 VDC.
DAC resolution is 40mV per count.
Pressure Limit Command DAC
The output of this DAC is the Pressure Limit Command signal
(PSILIM SELECTED). This signal is sent to the Multiplexer, the Pri-
mary Pressure Limit circuit, and the Overpressure Limit circuit.
Flow Rate Command DAC
The output of this DAC is the Flow Rate Command signal (FLOW
RATE SELECTED). This signal is sent to the Multiplexer, Flow Scale
Circuit and Power Drive Card (PDCI).
Mechanical Stop Position DAC
The output of this DAC is the Mechanical Stop Position signal
(MSPOSCMD). This signal is sent to the Multiplexer, and through
Unity Gain Buffer U26C, to the Mechanical Stop Drive Card.
Velocity Signal DAC
The output of this DAC is a digitized velocity signal, sent to the Imag-
ing System Interface Card. This signal is buffered, becoming the Flow
Profile Signal available through J106.
Multiplexer (MUX)
The multiplexer circuit is comprised of multiplexer U19, Quad latch
U20, and Unity-Gain buffer U26A. This circuit receives eight inputs,
and under timing control connects each input to the circuit output,
which is sent to the A/D converter.
TTL/CMOS Converter
U25 converts TTL logic (0/5V) to CMOS logic (0/15V).
10V Reference
Chip U28 de10 VDC as a reference voltage for the Plunger
and Mechanical Stop position feedback pots in the head. This voltage
is also applied to the multiplexer to be monitored by the CPU.
Pot Processor
This circuit, comprised of U27C and associated components, buffers
and filters the pot wiper signal. The pot signal is an input to the multi-
plexer and the error amplifier.
Primary Pressure
Limit Circuit
The Primary pressure limit circuit (U26B), compares the pressure limit
command to the actual pressure developed during an injection. The
output of this circuit is sent to three Pressure Limiting circuits to per-
form the following functions:
Summary of Contents for ProVis Mark V
Page 1: ...Service Manual KMP 870 98411 T 205 Rev D...
Page 2: ......
Page 3: ......
Page 4: ......
Page 15: ...Operating Guide 3 1 3 Operating Guide...
Page 16: ...Mark V ProVis Injection System 3 2...
Page 30: ...Mark V ProVis Injection System 7 4 Figure 7 1 Block Diagram Servo Control Card...
Page 31: ...Servo Control Card 7 5...
Page 37: ...Mark V ProVis Injection System 9 4 Figure 9 1 Block Diagram Control Panel Digital Logic...
Page 39: ...Mark V ProVis Injection System 9 6 Figure 9 2 Block Diagram Control Panel DC to DC Converter...
Page 49: ...Mark V ProVis Injection System 10 10 NOTES...
Page 60: ...Injector Head 13 7 Figure 13 1 Block Diagram Head Card...
Page 61: ...Mark V ProVis Injection System 13 8 NOTES...
Page 64: ...Imaging System Interface 14 3 Figure 14 1 Connector Locations...
Page 67: ...Mark V ProVis Injection System 14 6 Figure 14 2 Block Diagram Imaging System Interface...
Page 71: ...Mark V ProVis Injection System 14 10 NOTES...