M631 Precision RTD Simulator
MEATEST
67
Operation manual
9.
Technical data
Resistance range
:
16
… 400 k
SHORT, OPEN terminals
Pt sensor temperature simulation
:
-200.000
C … 850.000
C (-328
F … 1562
F)
Ni sensor temperature simulation
:
-60.000
C … 300.000
C (-76
F … 572
F)
Type of temperature sensors
:
Pt100 … Pt1000, Ni100 … Ni1000
Resolution
:
from 0.1 m
Pt temperature standards
:
IEC 751 (1,3850 for IPTS68)
(A=3.90802e-3, B=-5.80195e-7, C=-4.2735e-12)
IEC 751 (1,3851 for ITS90)
(A=3.9083e-3, B=-5.775e-7, C=-4.18301e-12)
1,3916
(A=3.9692e-3, B=-5.8495e-7, C=-4.2325e-12)
1,3926
(A=3.9848e-3, B=-5.870e-7, C=-4.0e-12)
Ni temperature standards
:
DIN 43760 (6180)
(A=5.485e-3, B=6.65e-6, C=2.805e-11, D=-2e-17)
Maximal dissipation power
:
0.25 W
Maximal current
:
0.5 A
Maximal voltage
:
200 Vpk
Reaction time *
:
6 ms
Terminals
:
instrument terminals diameter 4mm, gold plated
Interface
:
RS232 (optionally IEEE488, USB, Ethernet)
Power supply
:
115/230 Vac, 45…65 Hz
Maximal power consumption
:
15 VA
Reference temperature
:
+
20
C … +26
C
Working temperature
:
+
5
C … +40
C
Storing temperature
:
-10
C … +50
C
Housing
:
metal sheet
Dimensions
:
W 390 mm, H 128 mm, D 310 mm
Weight
:
5.2 kg
Languages
:
English, German, French, Spanish, Russian, Czech
Isolation resistance between output terminals and housing :
> 2 G
(at 500V DC)
*
Reaction time means time interval between setting up value from front panel or receiving
command from remote control bus and settling set-up value on output terminals. Value is valid
for FAST switching mode.
Notes:
Only data shown with tolerance or with band of limits are tested. All other values have
informative character.