Process of calibration consists of measuring partial resistances and writting their actual values
into the RTD simulator:
Set the first calibration point (resistance element). Use display softkeys “Previous” and
“Next” to select the element.
Measure resistance of the selected element. Use ohm-meter with appropriate accuracy in 4-
wire connection mode.
Figure 23 Calibration point screen
Using cursor keys ▲,▼, ◄, ► adjust resistance value in M631 according to the ohm
meter.
Confirm new calibration value by pressing “Save” softkey.
Repeat above described procedure for all resistance elements.
Change password
In this menu you can change the password of the calibration mode. Default password is “2”.