MEATEST
M631 Precision RTD Simulator
Operation manual
46
[:SOURce]:TIMing:PRESet:RAPPend
Syntax:
[:SOURce]:TIMing:PRESet:RAPPend <SPD>
Description:
This command appends new record at the end of timing table. Data will be saved to non-
volatile memory on TIM:PRES:SAVE command.
Parameters:
<SPD>
Quoted string representing amplitude. The amplitude consists of two
float numeric fields separated by comma. The first one represents
timing interval in seconds and the second one amplitude in Ohms.
Remarks:
Overlapped command
Example:
TIM:PRES:RAPP "0.5,220.0"
[:SOURce]:TIMing:PRESet:RCOunt?
Syntax:
[:SOURce]:TIMing:PRESet:RCOunt?
Description:
This commands returns actual number of records in timing table.
Parameters:
<DNPD>
Integer value representing number of records
Remarks:
Overlapped command
Example:
TIM:PRES:RCO? Response: 6
[:SOURce]:TIMing:PRESet:ROW<IND_ROW>:AMPLitude
Syntax:
[:SOURce]:TIMing:PRESet:ROW<IND_ROW>:AMPLitude <SPD>
[:SOURce]:TIMing:PRESet:ROW<IND_ROW>:AMPLitude?
Description:
This command sets / retrieves selected row in timing table. Data will be saved to non-
volatile memory on TIM:PRES:SAVE command.
Parameters:
<IND_ROW>
Range 1 ... Row count (1 - if omitted)
<SPD>
Quoted string representing amplitude. The amplitude consists of two
float numeric fields separated by comma. The first one represents
timing interval in seconds and the second one amplitude in Ohms.
Remarks:
Overlapped command
Example:
TIM:PRES:ROW1:AMPL "0.5,220.0"
TIM:PRES:ROW1:AMPL? Response: " 5.000000E-01,2.02"