![Meatest M631 Operation Manual Download Page 10](http://html.mh-extra.com/html/meatest/m631/m631_operation-manual_1758445010.webp)
MEATEST
M631 Precision RTD Simulator
Operation manual
10
4.
Operation
4.1.
Connection and disconnection of output terminals
Set value is connected (disconnected) to the output terminals after pressing OPER key.
Connected output terminals are indicated by the lit LED in the key.
Disconnected output terminals can be used for “Open terminals” simulation. “Short circuit” is
simulated after pressing SHORT key. Active SHORT key (LED in the key is ON) replaces the main
value with the short circuit. Also the short circuit must be connected to the output terminals by the
OPER key.
4.2.
Wires connection
Output resistance is available on R output terminals. Available is 2, 3 and 4-wire connection.
Both sides (red and grey) are floating up to 500Vpk against the case (PE).
Ground terminal is connected to the metal case and to the protective earth (PE).
4.3.
Setting the function
Function can be changed after pressing „Function“ softkey. New function is selected using
cursor keys ▲,▼ or display softkeys. Selection must be confirmed by pressing SELECT key or „OK“
softkey.
Device supports following functions:
Resistance
Offers direct setting of exact resistance value.
Figure 5 Resistance screen
Editable parameters:
Resistance value:
16 Ω … 400 kΩ
Switching mode:
FAST, SMOOTH, VIA OPEN, VIA SHORT
Summary of Contents for M631
Page 1: ...M631 Precision RTD Simulator Operation manual ...
Page 2: ......