M631 Precision RTD Simulator
MEATEST
61
Operation manual
:SYSTem:LOCal
Syntax:
:SYSTem:LOCal
Description:
This command places device in the LOCAL mode and unlocks all keys on front panel of
the device. The Command is valid only for RS232, LAN and USB interfaces. The device
will not respond to commands in LOCAL mode.
Parameters:
None
Remarks:
Overlapped command
Example:
SYST:LOC
:SYSTem:PRESet
Syntax:
:SYSTem:PRESet
Description:
This command will set all device (functions) parameters to default. The same as *RST
without an actual restart.
Parameters:
None
Remarks:
Overlapped command
Example:
SYST:PRES
:SYSTem:REMote
Syntax:
:SYSTem:REMote
Description:
This command places device in the REMOTE mode and locks all keys but LOCAL key.
The Command is valid only for RS232, LAN and USB interfaces. The device will not
respond to any other command until is in REMOTE mode.
Parameters:
None
Remarks:
Overlapped command
Example:
SYST:REM
:SYSTem:RWLock
Syntax:
:SYSTem:RWLock
Description:
This command places device in the REMOTE mode and locks all keys including LOCAL
key. The Command is valid only for RS232, LAN, USB interfaces. The device will not
respond to any other command until is in REMOTE mode.
Parameters:
None
Remarks:
Overlapped command
Example:
SYST:RWL
:SYSTem:TIME
Syntax:
Summary of Contents for M631