
MEATEST
M-140 Multifunction Calibrator
Operation manual v46
91
OUTP :ISEL (?) <CPD> { HIGH | HI50 }
This command activates or deactivates the 1000A current range (using a 50-turn coil).
•
HIGH - deactivates the 50-turn coil
•
HI50 - activates the 50-turn coil (up to 1000A range)
If query is sent, M140 returns HIGH if the 50-turn coil is deactivated or HI50 if it is activated.
Example:
OUTP :ISEL HI50 <cr> activates the 50-turn coil
OUTP :ISEL ? <cr> the calibrator returns HIGH or HI50
SOURce subsystem
This subsystem allows to control the individual functions of M140 calibrator.
Keyword
Parameters
[SOURce]
: FUNCtion
[: SHAPe] (?)
<CPD> { DC | SINusoid | PULPositive | PULSymmetrical |
PULNegative | RMPA | RMPB | TRIangle | LIMSinusoid |
PWMPositive | PWMSymmetrical | PWMNegative | .
SQUare }
: VOLTage
[: LEVEl]
[: IMMediate]
[: AMPLitude] (?)
<DNPD>
: CURRent
[: LEVEl]
[: IMMediate]
[: AMPLitude] (?)
<DNPD>
: RESistance
[: LEVEl]
[: IMMediate]
[: AMPLitude] (?)
<DNPD>
: CAPacitance
[: LEVEl]
[: IMMediate]
[: AMPLitude] (?)
<DNPD>
: POWEr
[: LEVEl]
[: IMMediate]
[: AMPLitude] (?)
<DNPD>
: PHASe
: UNITS (?)
<CPD> { DEG | COS }
[: ADJust] (?)
<DNPD>
: VOLTage
[: LEVEl]
[: IMMediate]
[: AMPLitude] (?)
<DNPD>
: CURRent
[: LEVEl]
[: IMMediate]
[: AMPLitude] (?)
<DNPD>
: EARTh
: VOLTage (?)
<CPD> { ON | OFF | 0 | 1 }
: CURRent (?)
<CPD> { ON | OFF | 0 | 1 }