M-140 Multifunction Calibrator
MEATEST
62
Operation manual v46
CAPACITANCE function
nominal value [F]
set limits [F]
range [F]
1 n
5 p
900 p – 2.5 n
900 p
5 p
900 p – 2.5 n
2.5 n
5 p
900 p – 2.5 n
5 n
5 p
5 n - 10 n
2.5 n
5 p
5 n - 10 n
10 n
10 p
5 n - 10 n
20 n
10 p
10 n - 50 n
10 n
10 p
10 n - 50 n
50 n
50 p
10 n - 50 n
100 n
50 p
50 n - 250 n
50 n
50 p
50 n - 250 n
250 n
250 p
50 n - 250 n
500 n
250 p
250 n – 1 u
250 n
250 p
250 n – 1 u
1 u
1 n
250 n – 1 u
2 u
2 n
1 u – 2.5 u
1 u
2 n
1 u – 2.5 u
2.5 u
5 n
1 u – 2.5 u
5 u
5 n
2.5 u – 10 u
2.5 u
5 n
2.5 u – 10 u
10 u
20 n
2.5 u – 10 u
20 u
20 n
10 u – 50 u
10 u
20 n
10 u – 50 u
50 u
200 n
10 u – 50 u
Table C
FREQUENCY function
amplitude calibration
nominal value [V]
set limits [V]
range [V]
note
0.0
10 u
--
zero calibration
10
1 m
10
slope calibration
Table F
ANALOG INPUT multimeter function voltage, current, resistance calibration
nominal value [-]
set limits [-]
range [-]
note
cable adapter
0 V
50 u
10 V
zero calibration
Option 40
10 V
200 u
10 V
slope calibration
Option 40
0 mA
50 nA
20 mA
zero calibration
Option 40
19 mA
500 nA
20 mA
slope calibration
Option 40
0
5 m
200
zero calibration
Option 60
100
5 m
200
slope calibration
Option 60
0
50 m
2000
zero calibration
Option 60
1000
50 m
2 k
slope calibration
Option 60
0 mV
2 u
20 mV
zero calibration
Option 60
19 mV
2 u
20 mV
slope calibration
Option 60
0 mV
7 u
200 mV
zero calibration
Option 60
190 mV
7 u
200 mV
slope calibration
Option 60
0 mV
50 u
2 V
zero calibration
Option 60
1.9 V
50 u
2 V
slope calibration
Option 60
Table Multimetr
For calibration of ranges 10 V and 20 mA cable adapter Option 40 must be used. For calibration of resistance
mVDC ranges cable adapter Option 60 is necessary.