M-140 Multifunction Calibrator
MEATEST
100
Operation manual v46
Example: to set the temperature of cold end of thermocouple to 25
C:
MEAS :CONF :TEMP :THER :RJUN 25 <cr>
If query is sent, M140 returns the set temperature using standard exponential format. Example: 20.5
C is returned
as 2.001.
MEAS :CONF :SGS :VOLT (?) <DNPD>
This command selects the function SGS of internal multimeter, sets the supply voltage of sensor and switches the
multimeter on.
<DNPD>
Represents the supply voltage in V. 0V to 20V can be set.
Example: to set the multimeter to SGS measurement mode and power supply of sensor to 10V:
:MEAS :CONF :SGS :VOLT 10 <cr>
If query is sent, M140 returns the supply voltage of SGS using standard exponential format. Example: 15V is
returned as 1.001.
MEAS :CONF :OFF
This command switches the multimeter off.
TESTer subsystem
This subsystem allows to control features of M140 calibrator when used as tester. It starts selected test procedure
and reads its result.
Keyword
Parameters
TESTer
: RUN
<DNPD>
: RESUlt ?
TEST :RUN <DNPD>
This command starts selected procedure. Number 1 to 10 can be set.
<DNPD>
Represents number of procedure.
Example: to run test procedure 3 :
TEST :RUN 3 <cr>
TEST :RESU ?
M140 returns the result of selected procedure { PASS | FAIL | RUN }.
String „RUN“ is returned if selected procedure isn’t finished.
I/D (instrument identification)