M-140 Multifunction Calibrator
MEATEST
86
Operation manual v46
AC voltage test
Function
Range
Value (V)
Frequency (Hz) Deviation
allowed(%value)
V-AC
20 mV
0.019
1000
0.358
V-AC
200 mV
0.19
1000
0.142
V-AC
2.0 V
1.9
1000
0.030
V-AC
20.0 V
19.0
50
0.030
V-AC
20.0 V
19.0
120
0.030
V-AC
20.0 V
19.0
10000
0.030
V-AC
20.0 V
19.0
20000
0.082
V-AC
20.0 V
19.0
50000
0.082
V-AC
240.0 V
190.0
1000
0.036
V-AC
1000 V
750.0
120
0.057
Table IV
200 mA DC Basic range with linearity test
Function
Range
Value (A)
Frequency (Hz) Deviation
allowed(%value)
A-DC
200.0 mA
0.02
0.040
A-DC
200.0 mA
0.04
0.025
A-DC
200.0 mA
0.06
0.020
A-DC
200.0 mA
0.08
0.018
A-DC
200.0 mA
0.10
0.016
A-DC
200.0 mA
0.12
0.015
A-DC
200.0 mA
0.14
0.014
A-DC
200.0 mA
0.16
0.014
A-DC
200.0 mA
0.18
0.013
A-DC
200.0 mA
0.19
0.013
A-DC
200.0 mA
-0.02
0.040
A-DC
200.0 mA
-0.04
0.025
A-DC
200.0 mA
-0.06
0.020
A-DC
200.0 mA
-0.08
0.018
A-DC
200.0 mA
-0.10
0.016
A-DC
200.0 mA
-0.12
0.015
A-DC
200.0 mA
-0.14
0.014
A-DC
200.0 mA
-0.16
0.014
A-DC
200.0 mA
-0.18
0.013
A-DC
200.0 mA
-0.19
0.013
Table V
DC current test
Function
Range
Value (A)
Frequency (Hz) Deviation
allowed(%value)
A-DC
200.0 uA
0.00019
0.061
A-DC
200.0 uA
-0.00019
0.061
A-DC
2.0 mA
0.0019
0.025
A-DC
2.0 mA
-0.0019
0.025
A-DC
20.0 mA
0.019
0.013
A-DC
20.0 mA
-0.019
0.013
Table VI