MEATEST
M-140 Multifunction Calibrator
Operation manual v46
87
AC current test
Function
Range
Value (A)
Frequency (Hz) Deviation
allowed(%value)
A-AC
200.0 uA
0.00019
60
0.161
A-AC
2.0 mA
0.0019
60
0.081
A-AC
20.0 mA
0.019
60
0.055
A-AC
20.0 mA
0.019
120
0.055
A-AC
20.0 mA
0.019
1000
0.055
A-AC
20.0 mA
0.019
1000
0.055
A-AC
200.0 mA
0.19
60
0.055
Table VII
AC/DC high current test
Function
Range
Value (A)
Frequency (Hz) Deviation
allowed(%value)
A-DC
2.0 A
1.0
0.025
A-DC
2.0 A
-1.0
0.025
A-AC
2.0 A
1.0
60
0.060
A-DC
10.0 A
10.0
0.040
A-DC
10.0 A
-10.0
0.040
A-AC
10.0 A
10.0
60
0.160
Table VIII
AC/DC power test
Function
Range
Value (VA)
Frequency (Hz) Deviation
allowed(%value)
P 1
480W
480
60
0.07
P 0,5LA
480W
240
60
0.46
P 0,5LE
480W
240
60
0.46
P 1
2400W
2400
60
0.14
P 0,5LA
2400W
1200
60
0.77
P 0,5LE
2400W
1200
60
0.77
Table IX
Voltage = 240V
Resistance test
Function
Range
Value (Ohm)
Frequency (Hz) Deviation
allowed(%value)
O-4W
100.0 Ohm
10.0
DC
0.130
O-4W
100.0 Ohm
100.0
DC
0.015
O-4W
1000 Ohm
1000.0
DC
0.015
O-4W
10k Ohm
10000.0000
DC
0.015
O-4W
100k Ohm
100000.0000
DC
0.015
O-4W
1M Ohm
1.6
DC
0.050
O-4W
10M Ohm
1.7
DC
0.200
O-4W
50M Ohm
5.7
DC
0.500
Table X
Capacitance test
Function
Range
Value (F)
Frequency (Hz) Deviation
allowed(%value)
CAP
1 nF
1.00000e-09
1000
2.000
CAP
10 nF
1.000000e-8
1000
0.500
CAP
100 nF
1.000000e-7
1000
0.500
CAP
1 uF
1.000000e-6
500
0.500
CAP
10 uF
1.000000e-5
300
1.500
CAP
50 uF
5.000000e-5
300
2.000
Table XI