M-140 Multifunction Calibrator
Operation manual v46
113
The multimeter allows the measurement of small DC voltages in the range of 0 to 2 V. 140-41 cable adapter is
required. The signal to be measured connects to Hu/+mV and Lu/-mV terminals. Lu/-mV is the common terminal
of the multimeter. To activate the measurement, select mVDC function mode using METER button in the function
menu and press INPUT to measure the input value.
Measurement of resistance or temperature using resistance temperature sensors
Resistance can only be measured using four-wire connection using Opt. 60 or Opt. 140-41 cable adapter. Opt. 60
cable ends with four bananas, labeled Hi, Hu, Lu, Li. Their meaning is as follows:
•
Hi
current terminal H
•
Hu
voltage terminal H
•
Lu
voltage terminal L
•
Li
current terminal L
During the measurement of resistance or temperature using resistance temperature sensors, the rules applicable to
the four-wire connection must be observed.
Connection of Pt100 sensor using Opt. 60 cable:
Connection of Pt100 sensor using 140-41 cable adapter
Pt100
Hi
Hu
Lu
Li