M-140 Multifunction Calibrator
MEATEST
10
Operation manual v46
Pin
Label
Signal
Limitation
1
0V5MER common terminal of multimetr power supply source
2
GND
ground (protection earth)
3
SIMLI
RC simulator output, current terminal Li
Umax.= 10Vpp, Imax.=40mA
4
SIMLU
RC simulator output, voltage terminal Lu
Umax.= 10Vpp, Imax.=40mA
5
GND
ground (protection earth)
6
L
common terminal of multimeter input
7
-U
low output terminal for DC voltage range
8
-I
low output terminal for DC current range
9
NG2
sort function output, contact 2 of sort relay
Umax.=50Vpp, Imax.=100 mA
10
PTLI
resistance temperature sensor input terminal Li
Umax.= 10Vpp. R<2 k
11
PTLU
resistance temperature sensor input terminal Lu
input terminal L on ranges 20, 200, 2000 mV
Umax.= 10Vpp. R<2 k
12
TEST1
identification terminal of actually used adapter
13
TEST3
identification terminal of actually used adapter
14
0V5MER common terminal of multimeter power supply source
15
NC
not used
16
SIMHI
RC simulator output, current terminal Hi
Umax.= 10Vpp, Imax.=40mA
17
SIMHU
RC simulator output, voltage terminal Hu
Umax.= 10Vpp, Imax.=40mA
18
NC
not used
19
INP
multimeter input terminal for voltage/current ranges
Umax.=25 Vpp, Imax.=25 mA
20
+U
high output terminal for DC voltage range
Umax.=20 Vss
21
+I
high output terminal for DC current range
Imax.=25 mA
22
NG1
sort function output, contact 1 of relay
Umax.=50Vpp, Imax.=100 mA
23
PTHI
resistance temperature sensor input terminal Hi
Umax.= 10Vpp. R<2 k
24
PTHU
resistance temperature sensor input terminal Hu
input terminal H on ranges 20, 200, 2000 mV
Umax.= 10Vpp. R<2 k
25
TEST2
identification terminal of actually used adapter
TEST1
Functional inputs and outputs present at the connector can be best utilized using supplied cable adapters.
LCD display shows all information provided by the calibrator, e.g. set parameters of the signal, error messages,
setup information. The display is divided to several information sections.