M-140 Multifunction Calibrator
MEATEST
118
Operation manual v46
Function setup:
Calibrator
T RTD function
OUTPUT 140-41 ON
Multimeter
DCV or DCI depending on the output signal type of the unit
being tested
Adapter
Opt. 140-41
2. Calibration of industrial thermometers with thermocouple sensor and frequency output:
Function setup:
Calibrator
T TC function
OUTPUT 140-41 ON
Multimeter
F
Adapter
Opt. 140-41
3. Calibration of a single phase “electrical power/current loop” transducer
Function setup:
Calibrator
P function (voltage, current, power factor, frequency)
Multimeter
DCI
Adapter
Opt. 40
DUT
INPUT OUTPUT
Pt100
H L
DUT
INPUT OUTPUT
TC
H L
INPUT
OUTPUT
HU LU HI LI +I -I
DUT – power transducer
Opt. 40 adapter