Zeroing function can be used e.g. to compensate the voltage drops at the cables, to suppress any remanent value
of the signal etc.
Notes:
*
The current inputs of the multimeter are protected from current overloading by a fuse located in 140-41
cable adapter. The fuse can be replaced after turning the cap of the fuse holder. The procedure is
described in the adapter’s manual.
*
The multimeter can be used as a DC milivoltmeter with 20 mV, 200 mV, 2 V, 10 V ranges. The
appropriate input terminals are PTHU and PTLU.
*
Maximum allowed voltage on the input terminals against ground terminal is 20 V.
*
Frequency measurement is possible up to 15 kHz. Input signal must be within 0.2 to 5 V range.
squarewave or pulse shape of the input signal is expected.
Overloading can damage the multimeter.