Page 37
Control Commands
optoNCDT ILR 1181 / 1182
RM is only effective in DT mode.
It consists of three parameters which are separated by space (20
hex
).
X
Designates the number of preceding measuring values that will be evaluated in the case of non-con-
forming measurement. A maximum of ten preceding measured values can be evaluated.
Y
Defines the range of permissible values. If this range is exceeded in negative or positive direction, the
respective measuring value will be corrected accordingly.
Z
Stands for the number of values that are out of the permissible value range; in the event of out-of-toler-
ance values arriving in succession, the most recently corrected value will be included in the correction
process for the next out-of-tolerance value.
The maximum allowed number of out-of-tolerance values is 100.
Example:
- x = 3
- 2y = 0.1
- z = 1
9.83 (.387)
9.85
(.388)
9.88
(.389)
10 (.393)
9.89
(.389)
9.85
(.388)
9.86
(.388)
9.88
(.389)
9.75 (.384)
9.77
(.385)
9.70 (.382)
9.75 (.384)
9.80 (.386)
9.85 (.388)
9.90 (.39)
9.95 (.392)
10.00 (.393)
10.05 (.396)
1
2
3
4
5
6
7
8
9
10
Ist-Measurement
Korr. run
Limit
x
x
z
2y
Fig. 22 Correction of measuring value