Page 34
Control Commands
optoNCDT ILR 1181 / 1182
The measuring time setting option allows also the modifying of the measuring rate, for example, in order to
restrict the data volume or for synchronization purposes.
Measuring time can only be set as an approximate value, because the underlying principle of measurement
is subject to certain variances that cannot be accounted for:
DT measuring mode > measuring time approximately ST x 240 ms (except ST = 0)
DS measuring mode > measuring time approximately ST x 150 ms (except ST = 0)
Example:
The target distance is 25 m, but the target’s reflectance is not ideal. With a measuring time setting of ST 2,
E15 will be output following measurement. The user must increase the measuring time in this case!
i
One should work in DW or DX mode where stable measuring times are required.
7.3.5
SFx.x.....Display/Set Scale Factor
Standard setting: 1
SF multiplies a calculated distance value with a user-selectable factor for changes in resolution or outputs in
a different unit of measure. The scale factor may also be negative.
Scale factor
Resolution
Output
Unit of measure
SF1
1 mm
02.693
m
SF10
0.1 mm
26.931
dm
SF1.0936
0.01 yard
02.945
yard
SF3.28084
0.01 feet
08.835
feet
SF0.3937
1 inch
01.060
100 inch
SF-1
1 mm
-02.693
m
Fig. 20 Examples of scale factor
i
Following a change in the scale factor, the settings for digital and/ or analog output and offset must be
matched accordingly!