Page 48
Cleaning
optoNCDT ILR 1181 / 1182
E53
One or more parameters in the EEPROM not
set (Consequence: Division by 0)
1. Parameter SF examine (SF must be unequal 0)
2. Contact technical support --> reship the sensor
for repair.
E54
Hardware error (PLL)
Contact technical support --> reship the sensor for
repair.
E55
Hardware error
Contact technical support --> reship the sensor for
repair.
E61
Used parameter is inadmissible, invalid com-
mand sent
Check control software commands.
E62
1. Hardware error
2. wrong value in interface communication
(Parity error SIO)
Check external software parity setting.
E63
SIO overflow
Check time of emitted signals in application soft-
ware; integrate delay on transmission if necessary.
E64
Framing-Error SIO
Data format of the serial interface examine (8N1)
11.
Cleaning
Remove dust from optical surfaces (transmitter and receiver optics) with a blower brush.
Do not use cleaners that contain organic solvents, when wiping optical surfaces down
Contact the manufacturer in the case of stubborn contamination or soiling.
Avoid the use of any kind of solvents to clean the sensor.
> Damage of the sensor
Do not open the device. Do not loose any screw at the sensor
> Damage of the sensor