28
8 DETAILED PROCEDURES FOR NAVIGATING REVIEW, SET-UP, CALIBRATION
AND TEST MODE SUBMENUS
The following sections provide step-by-step procedures for navigating all the modes and submenus.
8.1 Review Mode
Review mode allows the settings of the detector to be reviewed safely without any changes being made.
Review mode consists of 7 submenus as shown in the table below:
Review
submenu
Icon
Settings Displayed
Software
SW
Revision and checksum
Alarms
Sensor cartridge and gas ID codes, Alarm 1 and alarm 2 configuration and set
points, time delay, latching/non latching, energized/de-energized
Faults
Configuration, Latching/non latching, energized/de-energized
Calibration
Days left to next calibration due and date of last calibration
Time/Date
timE
Current year, date and time
IP address
Auto address selection on/off, IP address, sub net mask values.
Event log
View the date, time and details of the last seven alarm, fault and maintenance
fault events
Table 11. Review mode submenus.
Review mode allows the settings of the detector to be reviewed safely without any changes being made. Review
mode consists of 7 submenus as shown in the table below:
To select review mode from normal operation, press the ‘
▲
’ up or ‘
▼
’ down button once. The
icon will be
displayed along with the first submenu icon. Press the ‘
▲
’ up or ‘
▼
’ down buttons to select a different submenu
icon.
8.1.1 Review Software ‘SW’
1. Select the software ‘SW’ submenu and press ‘
✓
’ to accept.
2. The software version number is displayed.
3. Press ‘
✓
’ to display the software checksum
4. Press ‘
✓
’ to return to step 1.
5. Repeat steps 1 through 4 to view the information again or press ‘
▲
’ up or ‘
▼
’ down to select another
submenu.
6. Press ‘X’ to return to normal operation