3-13
Masterflex REGLO Digital Pump Drive with Advanced Connectivity
Section 2: Basic Setup & Settings
Masterflex
Auto Start
The REGLO has an auto start that will resume pump operation when power is restored after a power outage.
To enable auto start:
1. Tap SETTINGS from any of the mode screens. The Settings screen will be displayed.
2. SCROLL through the Settings screen pages to locate AUTO START.
3. Tap the AUTO START toggle to select either ON or OFF.
Dispense & Batch Completion Alerts
When Sound Dispense is enabled a short beep will sound at the completion of each pump dispense. When
Sound Batch is enabled a single long beep will sound at the completion of each batch cycle.
To enable or disable the dispense completion alert:
1. Tap SETTINGS from any of the mode screens. The Settings screen will be displayed.
2. SCROLL through the Settings screen pages to locate SOUND DISPENSE.
3. Tap the SOUND DISPENSE toggle to select either ON or OFF.
To enable or disable the batch cycle completion alert:
1. Tap SETTINGS from any of the mode screens. The Settings screen will be displayed.
2. SCROLL through the Settings screen pages to locate SOUND BATCH.
3. Tap the SOUND BATCH toggle to select either ON or OFF.
Analog Output
The REGLO supports and controls analog output through the pump drive’s 25-pin connection (for further
1. Tap SETTINGS from any of the mode screens. The Settings screen will be displayed.
2. SCROLL through the Settings screen pages to locate ANALOG OUTPUT.
3. Tap ANALOG OUTPUT. The Analog Output screen will be displayed.
4. Tap the desired analog output type from the available list.
5. If using either ANALOG: VOLTAGE or ANALOG: CURRENT:
a. Tap EDIT . The Analog: Voltage or Analog: Current Edit screen will be displayed.
b. Select the desired current or voltage operating range from the available list.
c. Tap CONFIRM to save or CANCEL to discard changes.
The Analog Output screen will be
displayed.
6. Tap CONFIRM .