3-16
Masterflex REGLO Digital Pump Drive with Advanced Connectivity
Section 3: Operation
Masterflex
CONTINUOUS MODE
In Continuous Mode the pump will operate at a selected rpm and/or flow rate until stopped by the user.
Continuous Mode Run Screen
The Continuous Mode Run screen is accessed by selecting CONTINUOUS from the Mode Selection screen.
A. MODE NAME BANNER: Tap the Mode Name Banner to return to the Mode Selection screen.
B. CUMULATIVE VOLUME: Displays the current cumulative volume of fluid that has been pumped.
To
reset the cumulative volume:
Press and hold CUMULATIVE VOLUME RESET until the display
changes to zero.
C. FLOW RATE: Displays the current flow rate in the unit of measurement selected by the user. Tap to
access the Flow Rate screen.
D. SIZE: Tap to access the Tube Size screen.
E. CALIBRATE: Tap to access the Calibration screen (for further information see "
page 3-8). Once calibration has been completed for the current tube size Calibration Complete will
be displayed.
F. PRIME: Press and hold to prime the pump drive (for further information see "
G. START/STOP: Tap START to begin operation. Once operation has commenced STOP will be
displayed.
H. FLOW DIRECTION: Tap DIRECTION to select either clockwise
or counterclockwise
.
B. Cumulative Volume
G. Start/Stop
D. Tube Size
E. Calibrate
J. Settings
I. Screen Lock
C. Flow Rate
H. Flow Direction
A. Mode Name Banner
F. Prime