3-29
Masterflex REGLO Digital Pump Drive with Advanced Connectivity
Section 3: Operation
Masterflex
REMOTE MODE
!
CAUTION
: Power must be turned off before connecting the external remote control
cable to prevent damage to the drive.
The REGLO Digital Pump Drive features both digital remote control and monitoring through the
MasterflexLive™ website using an Ethernet or Wi-Fi connection and/or analog remote control and monitoring
via the pump drive’s DB-25 female connection port. See
for more information.
During remote operation:
• The Remote Mode Run screen will be displayed on the pump drive’s touchscreen.
• All touchscreen functions, with the exception of STOP , SCREEN LOCK
and EDIT will be
disabled.
Remote Mode Run Screen
The Remote Mode Run screen displays the currently selected operation dispense settings (Continuous, Time
or Volume) from an external remote control device. The Remote Mode Run screen is accessed by selecting
REMOTE from the Mode Selection screen.
A. MODE NAME BANNER: Displays the Mode Name as well as the currently selected remote type
(ANALOG CURRENT, ANALOG VOLTAGE or MASTERFLEXLIVE™). Tap the Mode Name Banner
to return to the Mode Selection screen.
B. EDIT: Tap to access the Remote Mode Edit screen. From the Remote Mode Edit screen adjustments can
be made to analog inputs and outputs for voltage and current (for further information see "
B. Edit
C. Stop
D. Screen Lock
A. Mode Name Banner