9. Tap CONFIRM to save or DELETE to discard changes.
NOTE:
• Once saved, new programs are added in alphabetical order to the bottom of the Mode Selection screen.
• If there are multiple programs tapping SCROLL will display additional pages.
Using Program Modes
Custom programs will appear in alphabetical order at the bottom of the Mode Selection screen. If there are
multiple programs tapping SCROLL will display additional pages.
General Preparation:
• Confirm the correct tubing has been selected and is properly loaded into the pump drive (for further
• Confirm all fluids and containers are ready.
• Confirm that the tubing has been calibrated by checking that Calibration Complete is displayed on
the Program Information screen (for further information see "
• If required, prime the pump before operation (for further information see "
page 3-7).
To use a program:
1. Select the desired program from the Mode Selection screen. The Program Run screen will be displayed.
2. Tap START to begin the pump drive operation using the displayed program settings.
3. If required, Tap PAUSE or STOP during operation.
NOTE:
The pump cannot be paused during
Continuous Mode operation.
NOTE:
• Tap INFORMATION to view the selected program's settings.
• Individual settings cannot be edited from the Program Run screen.
Editing a Program
To edit a program:
1. Tap EDIT located next to the desired program name on the Mode Selection screen (
NOTE:
If there
are multiple programs, tap SCROLL to display additional pages). The Program Edit screen will be
displayed.
2. Edit the settings for the selected mode as desired. For instructions on editing individual mode settings see:
• "
Adding a New Program: Continuous Mode
• "
Adding a New Program: Time Mode
• "