3-25
Masterflex REGLO Digital Pump Drive with Advanced Connectivity
Section 3: Operation
Masterflex
• If required, prime the pump before operation (for further information see "
page 3-7).
To operate the pump in Volume Mode:
1. Tap VOLUME from the Mode Selection screen. The Volume Mode Run screen will be displayed.
2. Tap EDIT to access the Volume Mode Edit screen.
3. Tap VOLUME. The Dispense Volume screen will be displayed.
a. Tap VOLUME and enter the desired volume using the onscreen keypad.
NOTE:
If the volume of
fluid is unknown the pump drive can record fluid dispense volumes (
for further information see
.
b. Tap RAMP UP to set the time the pump takes to reach full flow rate (between 0 and 60 sec-
onds) using the onscreen keypad.
c. Tap RAMP DOWN
to set the time the pump takes to stop at the end of each dispense opera-
tion (between 0 and 60 seconds) using the onscreen keypad.
d. Tap CONFIRM to save or CANCEL to discard changes and return to the Volume Mode
Edit screen.
4. Tap INTERVAL TIME to set the amount of time the pump will pause between each operation. The
Interval Time screen will be displayed.
NOTE:
If the interval time is set to 0 (zero) the pump drive will
require a start input (from either the touchscreen or a wired input) to start the next dispense.
a. Tap HR, MIN, SEC or TENTHS and enter the desired time using the onscreen keypad.
b. Tap CONFIRM to save or CANCEL to discard changes and return to the Volume Mode
Edit screen.
5. Tap FLOW RATE. The Flow Rate Edit screen will be displayed.
a. Tap UNITS to access the Flow Units screen and select the desired flow unit from the available list.
b. Tap CONFIRM to save or CANCEL to discard changes and return to the previous screen.
c. Tap FLOW and enter the desired flow rate using the onscreen keypad.
d. Tap CONFIRM to save or CANCEL to discard changes and return to the Volume Mode
Edit screen.
6. Tap SIZE. The Tube Size screen will be displayed.
a. Select the desired tube size from the available list or
SCROLL
and tap CUSTOM to enter a
custom tube size using the onscreen keypad.
b. Tap CONFIRM to save or CANCEL to discard changes and return to the Volume Mode
Edit screen.
7. Tap BATCH TOTAL to adjust the number of dispenses in each batch cycle. The Batch Total screen will
be displayed.
a. Enter the desired number of dispenses in a batch using the onscreen keypad.
NOTE:
If required,
tap INFINITE
∞
to select an infinite number of dispense cycles. If infinite is selected, the pump
will run continuously.