Masterflex REGLO Digital Pump Drive with Advanced Connectivity
Section 2: Basic Setup & Settings
Masterflex
8. Tap SIZE from the Continuous Mode screen or tap EDIT and then SIZE if either the Volume or
Time Mode screen is displayed. The Tube Size screen will be displayed.
a. Select the desired tube size from the available options or
SCROLL
and tap CUSTOM to enter
a custom tube size.
b. Tap CONFIRM to save or CANCEL to discard changes and return to previous screen.
9. Tap CALIBRATE . The Tube Calibration screen will be displayed.
10. Tap CALIBRATION VOLUME. The Calibration Volume screen will be displayed.
a. Enter the desired calibration volume using the onscreen keypad.
NOTE:
The default volume or
greater will ensure the best calibration.
b. Tap CONFIRM to save or CANCEL to discard changes and return to the previous screen.
11. Tap FLOW RATE. The Calibration Flow Rate screen will be displayed.
a. Enter the desired flow rate using the onscreen keypad.
b. Tap CONFIRM to save or CANCEL to discard changes and return to the previous screen.
NOTE:
The drive will adjust the displayed flow rate after calibration is complete.
12. Tap START to begin calibration.
NOTE:
Calibration progress will be displayed on the touchscreen.
13. When calibration is complete use the onscreen keypad to enter the measured volume of fluid from the
tube outlet container.
14. Tap
CONFIRM
to complete calibration and return to the previous mode screen. Calibration
Complete should now be displayed to indicate that calibration has been completed for the current
tube size, tube volume, and flow rate.
NOTE:
• The estimated accuracy for the selected calibration settings is shown in the Calibration Strength bar on
the Tube Calibration screen. Calibration accuracy can be improved by using a larger calibration volume
and/or altering the flow rate.
• Calibration settings are retained and transferred to other mode screens when entering or leaving the Tube
Calibration screen.
• If the drive is stopped during calibration the calibration will be discarded and the previous mode screen
will be displayed. Empty the container before restarting the procedure.
• Calibration time at the maximum allowable flow rate (default max flow rate) is approximately 10 seconds.
Calibration time at the minimum allowable flow rate (approximately 4% of the maximum flow rate) is 4
minutes. Select a custom tube size for other tubing sizes or lower flow rates (see step 8 above).
• Minimum and maximum flow rates will change after calibration due to a recalculation of the volume of
fluid per pump revolution.
• The best results are obtained after tubing has been run in the pump for approximately 20-30 minutes.
• Steps 9–14 can be repeated as necessary to optimize the accuracy of the tubing calibration.