3-31
Masterflex REGLO Digital Pump Drive with Advanced Connectivity
Section 3: Operation
Masterflex
Analog Input
!
CAUTION
: Power must be turned off before connecting the external remote
control cable to prevent damage to the drive.
The REGLO supports and controls analog input using suitable equipment through the pump drive’s DB-25
female connection port. Analog input can be used in place of, or in addition to, MasterflexLive™.
See also "
General Preparation:
• Confirm the correct tubing has been selected and is properly loaded into the pump drive (for further
• Confirm all fluids and containers are ready.
• Confirm that the tubing has been calibrated by checking that Calibration Complete is displayed on the
Remote Mode Edit screen (for further information see "
• If required, prime the pump before operation (for further information see "
page 3-7).
• Confirm that the pump drive is connected to the appropriate equipment through the drive’s DB-25
female connection port.
To operate analog control and monitoring:
1. Tap REMOTE from the Mode Selection screen. The Remote Mode Run screen will be displayed.
2. Tap EDIT . The Analog Input screen will be displayed.
3. Tap the desired analog input type from the available list.
4. If using either ANALOG: VOLTAGE or ANALOG: CURRENT:
a. Select the desired current or voltage operating range from the available list.
b. Tap CONFIRM to save or CANCEL to cancel and return to the previous screen.
NOTE:
• While the pump is operating the Remote Mode Run screen will be displayed on the pump drive’s
touchscreen and local touchscreen control will be unavailable, with the exception of STOP , SCREEN
LOCK
and EDIT .
• Tap STOP on the pump drive touchscreen to override the remote control device and immediately stop
the pump operation.
NOTE:
If locked, the screen will need to be unlocked before the pump operation
can be stopped.
•
To lock the screen:
Press and hold UNLOCKED until the icon changes to LOCKED and a red
border appears around the screen.
To unlock the screen:
Press and hold LOCKED until the icon
changes to UNLOCKED .
NOTE:
Stop is disabled when the screen is locked.
• See
for further information on using analog remote control and monitoring.