7. After completing a factory reset, LIFT ZERO SET screen
shows automatically. Press
OK
to accept this automatic
setting.
Note:
For eliminating accumulated position error of lifting column, adjust lifting column
zero position after performing a factory reset.
Then the PFD Home Position adjustment MUST be performed again. Refer to
section 8.7, Setting the PFD Home Position, on page 35.
8.13.
Servo Pack Parameter Information
1. To get or set parameters of Servo Pack from the standby
condition, press and hold the
MENU
key.
2. Press
UP
or
DOWN
key to select SET SERVO menu in the
SETUP screen.
3. Press
ENTER
key to set parameters of Servo Pack.
4. Press
UP
or
DOWN
key to change values of ADDRESS or
DATA
5. Press
LEFT
or
RIGHT
key to select ADDRESS, DATA, EXIT,
RD, or WR.
6. Press and hold the
ENTER
key on selecting
EXIT
to return to
the SETUP screen.
7. Press and hold the
ENTER
key on selecting
RD
to read data
of the defined address from Servo Pack.
8. Press and hold the
ENTER
key on selecting
WR
to write data
of the defined address into Servo Pack.
8.14.
System Log Information
Note:
The system can store up to 2048 unique operation logs referenced by date and
time. These can assist with system troubleshooting should a problem occur.
1. To get system information from the standby condition, press
the
MENU
key more than 1 second.
2. Press
UP
or
DOWN
key to select LOG INFO menu in the
SETUP screen.
3. Press
ENTER
key to get and see log data.