Rad-97
Chapter 4: Operation
www.masimo.com
63
Masimo
Options
Description
Alarm
Priority
Factory
Default
Settings
User
Configurable
Settings
Adaptive
Threshold
Alarm (ATA)
ATA establishes patient-specific limit
thresholds based upon the baseline
value of the parameter.
See
Adaptive Threshold Alarm (ATA)
Feature
on page 127.
NA
Off
Off or On
Additional Settings for SpO2
From the
Additional Settings
screen, change any of the following options:
Options
Description
Factory
Default
Settings
User Configurable
Settings
Averaging
Time*
The length of time over which the
system calculates the average of all
data points.
8 seconds**
2-4, 4-6, 8, 10, 12, 14,
or 16 seconds***
FastSat
See
FastSat Overview
on page 63.
Off
Off or On
* With FastSat the averaging time is dependent on the input signal.
** Defaults to 2-4 seconds when in Sleep Study mode. See
Sleep Study
on page 90.
*** For the 2 and 4 second settings the averaging time may range from 2-4 and 4-6 seconds,
respectively.
FastSat Overview
FastSat enables rapid tracking of arterial oxygen saturation changes. Arterial oxygen
saturation data is averaged using pulse oximeter averaging algorithms to smooth the trend.
When Rad-97 is set to FastSat
On
, the averaging algorithm evaluates all saturation values,
providing an averaged saturation value that is a better representation of the patient’s current
oxygenation status. With FastSat set to On, the averaging time is dependent on the input
signal.