Chapter 9: Alarms and Messages
www.masimo.com
128
Masimo
Note:
ATA does not allow a Rapid Desat setting of 0%.
When ATA is turned
Off
, the device uses the standard alarm limits and standard alarm delays.
See
About Alarms
on page 124
See
SpO2 Alarms
on page 62
3D Alarms
3D Alarms, accessible from the Main Menu, include the following:
Desat Index
on page 129
About Desat Index
on page 128
Pi Delta
on page 130
About Pi Delta
on page 129
About Desat Index
The 3D Desat Index Alarm allows a clinician to request audible and visual alarms if a patient
experiences a specified number of desaturations beyond a defined level from the patient's
baseline saturation over a specific period of time.
Traditional high and low SpO
2
alarm limits alert clinicians to saturation levels that exceed
user-selected thresholds. These thresholds are typically established to detect significant
changes from patients’ baseline saturation levels. However, in select patient populations,
substantial desaturation events that remain above a typical low alarm limit threshold may be
preceded by a cycle of smaller transient desaturations over a limited period of time. The
ability to alert clinicians when a cycle of smaller transient desaturations occur may provide
an earlier indication of a potential significant decline in patient status, allowing for more
focused monitoring and/or a change in treatment.