Chapter 10: Troubleshooting
www.masimo.com
140
Masimo
Troubleshooting Rad-97
The following section lists possible Rad-97 symptoms, potential causes, and next steps.
For more information, see
Chapter 9: Alarms and Messages
on page 123.
Symptom
Potential Causes
Next Steps
Device does not turn on
•
Depleted Battery.
•
Internal failure.
•
Check AC Power
connection.
•
Contact Masimo
Service. See
Contacting Masimo
on
page 173.
System failure technical alarm
active
•
Internal failure.
•
Turn Rad-97 Off and
On.
•
Contact Masimo
service. See
Contacting
Masimo
on page 173.
Speaker does not work
•
Device audible settings
may be incorrect.
•
Internal failure.
•
Turn Rad-97 Off and
On.
•
Check that
Alarms
and
Sounds
have not been
silenced.
•
Check that
Alarms
and
Sounds
volumes
settings.
•
Check the device is not
in
All Mute
.
•
Check that the device
speaker is not being
muffled.
•
Contact Masimo
service. See
Contacting
Masimo
on page 173.