Chapter 4: Operation
www.masimo.com
85
Masimo
Calibration for NIBP
The
Calibration
option on the
NIBP
menu allows a qualified service professional to access
calibration settings and tools for the NIBP module. For more information, see
Chapter 12:
Service and Maintenance
on page 167.
Note:
This section is provided as a reference and intended for qualified service professionals
only.
Sounds
Use the
Sounds
screen to control the volume of sounds and duration of audio pause on
Rad-97. Users can also access the
Sounds
screen by pressing the
Sounds
icon on the Status
Bar. See
About the Status Bar
on page 49.
Option
Description
Factory
Default
Settings
User Configurable
Settings
Alarm
Volume
Sets the alarm volume level.
4 (Highest
volume)
1 to 4 - Slide towards the
left to decrease volume
and to silence.
Pulse Tone
Volume
Sets the pulse tone volume level.
3
0 to 4 - Slide towards the
left to decrease volume
and to silence.
Audio Pause
Duration
Sets the length of time that the
audible alarm remains silenced,
when Audio Pause is enabled. See
Audio Pause
on page 126.
2 minutes
1, 2, or 3 minutes,
Permanent *,**, or
Permanent with
Reminder *,***.
SmartTone
Allows the audible pulse to continue
to beep when the pleth graph shows
signs of motion.
Off
On or Off
* Requires user to have All Mute Enabled turned on in the
Access Control
menu. See
Access
Control
on page 94.
** If
Permanent
is selected, there will be no audible alarms, but visual alarms will still display.