Rad-97
Chapter 9: Alarms and Messages
www.masimo.com
124
Masimo
Alarm Source/Example
Explanation
System Level:
The example
shown here is a "No Cable
Connected" alarm. Note that
the border of the entire Rad-97
display is illuminated, and the
explanation of the alarm is
shown in the Status Bar (No
Cable Connected).
About Alarms
The
Alarm Silence
icon is an indicator as well as a functional button. It always indicates the
presence of alarms, and it can be used to temporarily suspend audible alarms for a
pre-configured amount of time (Silence Duration).
Silence Duration configurations vary across different parameters and measurements. For
more information about Silence Duration, refer to
rainbow Parameter Settings
on page 59.
Icon
Appearance
Description
Visual
Alarms
There are currently no active alarms, and no alarms have been
silenced.
No
There are currently no active alarms, but at least one alarm
has been and is still silenced.
No
There is currently at least one active alarm that has
not
been
silenced.
Yes
There is currently at least one active alarm, but all active
alarms are silenced.
Yes