Rad-97
Chapter 6: Noninvasive Blood Pressure (NIBP)
www.masimo.com
111
Masimo
To ensure the correct cuff size
Wrap the cuff around the arm.
The index line
should align within the cuff range markings
.
If the index line does not fit within the range markings, select a larger or smaller cuff.
To place cuff on the measurement site
Wrap the cuff around the non-dominant arm, making sure that the Artery Marker is aligned
over the brachial artery as shown in the image below. If possible, do not wrap the cuff over the
patient’s clothing. The cuff should fit snugly around the patient’s arm for maximum
oscillometric signal quality. The lower edge of the cuff should be located 2 cm above the
antecubital fossa (interior bend of the elbow).
Ensure that the air hose from the monitor to the cuff is not compressed, crimped, or damaged.