Rad-97
Chapter 4: Operation
www.masimo.com
70
Masimo
Note that Rad-97 can monitor RRa or RRp but not both simultaneously.
RRp is active when the following conditions have all been met:
•
RRp is installed on the Rad-97.
•
No dual rainbow cable is connected.
•
A pulse oximetry or pulse CO-Oximetry sensor is connected.
•
The optical sensor must support RRp.
From the
RR Settings
screen, access any of the following screens:
RRp Alarms
on page 70
Additional Settings for RRp
on page 71
RRp Alarms
From the
Alarms
screen, change any of the following options:
Options
Description
Alarm
Priority
Factory
Default
Settings
User Configurable
Settings
High
Limit
High Limit is the upper
threshold that triggers an
alarm.
High
30 breaths
per minute
6 to 69 breaths per
minute in steps of 1
breath per minute, or Off
Low Limit Low Limit is the lower
threshold that triggers an
alarm.
High
6 breaths
per minute
Off, or 5 to 68 breaths
per minute in steps of 1
breath per minute
Silence
Duration
Sets the amount of time
that the alarm is silenced.
NA
2 minutes
30 seconds or 1, 2, or 5
minutes
Alarm
Delay
When an alarm condition is
met, this feature delays the
audible part of an alarm.
NA
30 seconds 0, 10, 15, 30, or 60
seconds