Chapter 4: Operation
www.masimo.com
62
Masimo
SpO2 Settings
Allows access to any of the following options:
SpO2 Alarms
on page 62
Additional Settings for SpO2
on page 63
In Vivo for SpO2
on page 64
Trends
on page 97
About Parameter Information
on page 78
About Desat Index
on page 128
SpO2 Alarms
From the
Alarms
screen, change any of the following options:
Options
Description
Alarm
Priority
Factory
Default
Settings
User
Configurable
Settings
High Limit High Limit is the upper threshold that
triggers an alarm.
Medium Off
2% to 99% in
steps of 1%, or
Off
When set to
Off, alarm is
disabled
Low Limit
Low Limit is the lower threshold that
triggers an alarm.
High
88%
1% to 98% in
steps of 1%
Rapid Desat
Sets the Rapid Desat limit threshold
to the selected amount below the Low
Alarm Limit. When an SpO
2
value falls
below the Rapid Desat limit the audio
and visual alarms are immediately
triggered without respect to alarm
delay.
NA
-10%
Off, -5%, or
-10%
Alarm Delay
When an alarm condition is met, this
feature delays the audible part of an
alarm.
NA
15
seconds
0, 5, 10, or 15
seconds
Silence
Duration
Sets the amount of time that the
alarm is silenced.
NA
2 minutes 30 sec, 1 or 2
minutes